FluoroMax-3 v. 3.1 (7 Mar 2006)
Producing Correction Factors
8-11
Note:
The signal detector, S, uses only
the
mcorrect
file; the reference de-
tector, R, uses only the
xcorrect
file.
The
Preferences
area appears.
5
Choose the Instrument Correction Files icon.
The
Instrument Correction Files
area appears.
6
If there are no active fields in the Instrument
Correction Files area, click Insert.
7
In sequence,
a
Choose the
Detector
column, and select the appropriate detector from the drop-down list,
b
The
Mono
column, and the correct monochromator from the drop-down
list,
c
The
Grating
column, the correct grating from the drop-down list,