FluoroMax-3 v. 3.1 (7 Mar 2006)
Components & Accessories
12-3
Note:
If uncertain of the absorption range, enter a range of 250 nm to
400 nm, and scan the excitation spectrometer. Detailed instructions
for setting parameters are contained in the on-line help. The S/R ac-
quisition mode compensates for fluctuations in lamp intensity, and
wavelength dependency of the lamp and excitation spectrometer.
Model 1940 Absorption/Transmission
Accessory
The Model 1940 Absorption/Transmission
Accessory slightly displaces the sample
from its normal position and directs the
transmitted light into the collection optics
with a mirror mounted at 45°.
Installation
1
Remove the sample
holder currently in place.
2
Position the Model 1940 on the posts.
3
Tighten the two thumbscrews.
Before transmission or absorption spectra are acquired, the emission and reference sig-
nals must be optimized.
Signal optimization
1
Place the cuvette containing the blank in the
sample holder.
2
Set the emission monochromator position to 0
nm.
3
Select acquisition mode
S/R
.
4
Run an excitation scan over the absorption
range of the sample.
Summary of Contents for FluoroMax-3
Page 5: ...FluoroMax 3 v 3 1 3 Mar 2006 v 15 Declaration of Conformity 15 1 16 Index 16 1...
Page 6: ...FluoroMax 3 v 3 1 3 Mar 2006 vi...
Page 18: ...FluoroMax 3 v 3 1 3 Mar 2006 Introduction 0 12...
Page 36: ...FluoroMax 3 v 3 22 Feb 2005 System Description 2 8...
Page 96: ...FluoroMax 3 v 3 1 3 Mar 2006 Maintenance 6 12...
Page 156: ...FluoroMax 3 v 3 1 6 Mar 2006 Automated Polarizers 10 22...
Page 208: ...FluoroMax 3 v 3 1 6 Mar 2006 Bibliography 14 4...
Page 210: ...FluoroMax 3 v 3 1 6 Mar 2006 Declaration of Conformity 15 2...