FluoroMax-3 v. 3.1 (3 Mar 2006)
Optimizing Data
5-15
Note:
Before applying correction factors, HORIBA Jobin Yvon Inc.
recommends subtracting the dark counts, and the spectrum of the
blank, from the data. See the on-line FluorEssence™ help files for
specific instructions.
Caution:
The
mcorrect
,
xcorrect
, and
tcorrect
files are custom-generated at the factory for each
particular instrument, and cannot be swapped.
During acquisition
Data can be acquired either as raw data or as corrected data. A spectrum composed of
raw data exhibits the effects of system parameters, while a corrected spectrum displays
only the properties related to the sample.
1
Be sure the instrument configuration has a lay-
out that includes a correction file associated with
the appropriate detector.
a
Choose
Collect
in the toolbar.
A drop-down menu appears.
b
Choose
Advanced Setup
, then
System Configuration
.
The
System Configuration
window appears.