FluoroMax-3 v. 3.1 (7 Mar 2006)
Components & Accessories
12-5
Note:
If the detectors are
saturated, close the slits a
little or install a neutral-
density filter.
a
Acquire a synchronous scan
with an offset of 0 nm over the
absorption range of the blank.
b
Acquire a synchronous scan
with an offset of 0 nm over the
absorption range of the sam-
ple.
11
Calculate the absorption.
The absorption spectrum can be determined from the following equation:
A
= log(
S
/
R
)
blank
– log(
S
/
R
)
sample
Summary of Contents for FluoroMax-3