FluoroMax-3 v. 3.1 (6 Mar 2006)
Automated Polarizers
10-9
Note:
Do not
check the Re-
m
easure Anisot-
ropy Only
check-
box.
Warning:
Read the Materials Safety
Data Sheets (MSDS) before using
colloidal silica or glycogen.
Choose various options:
•
Subtract
Dark
(recommended)
•
Reset to Mechanical Zero
—only if the polarizers
are definitely miscalibrated. This deletes the
previous calibration.
6
Place the LUDOX
®
or
glycogen in the sample
holder.
7
Click
the
Con-
tinue button.
The system rotates through the polarizers as shown in the checklist on the win-
dow. As each phase is completed, the checkboxes are updated.
If the sample is too
concentrated or dilute, the
software prompts you to
correct this.