FluoroMax-3 v. 3.1 (7 Mar 2006)
Components & Accessories
12-16
Warning
:
Always read the Material Safety Data
Sheet (MSDS) to understand the hazards of
handling your sample.
Model 1933 Solid Sample Holder
The Model 1933 Solid Sample Holder is designed for samples such as thin films, pow-
ders, pellets, microscope slides, and fibers. The holder consists of a base with a dial in-
dicating angle of rotation, upon which a bracket, a spring clip, and a sample block rest.
Model 1933 Solid Sample Holder (with sample block nearby).
Installation
1
Remove the present holder.
2
Position the base on the posts.
3
Tighten the two thumbscrews.
For pellets, crystals, creams, gels, powders, and
similar materials:
1
Fill the well of the block.
2
Place a quartz coverslip or Teflon
®
film over the
well.
This holds the sample in place when vertically positioned.
3
Carefully insert the block between the bracket
and spring clip, so that the sample is
perpendicular to the excitation light.