FluoroMax-3 v. 3.1 (3 Mar 2006)
Maintenance
6-8
Is the signal
within specifica-
tions?
Yes
No further adjust-
ment is necessary.
Continue to adjust the
screws, to improve the
water Raman signal.
No
Replace the lamp-housing cover. Replace
the FluoroMax
®
-3 cover.
STOP.
The FluoroMax
®
-3 is ready for operation.
After repeated ad-
justments, is the
signal still low?
No
Yes
Go to Chapter 7: Troubleshooting.
6
Acquire another water Raman scan.
a
Use the same parameters as in step 3.
b
Note the peak intensity.
c
Use the flowchart below.