FluoroMax-3 v. 3.1 (3 Mar 2006)
Maintenance
6-10
tion-modules, and optional trigger input from a TRIG-15/25 accessory. The board has
an internal ±15 V DC supply.
Monochromator drives and accessory controllers
The monochromator drives, drive electronics, and accessory electronics are found on
the underside of the optical platform. Electrical connections for the emission detector
and the basic system-electronics for AC-power conversion are also underneath the in-
strument.
To gain access to the drive
electronics, remove the
bottom cover of the
instrument. Remove all
screws on the bottom cover
except those that attach the
dust screens for the exhaust
fans. The leveling feet do
not need to be removed
before detaching the bottom
cover.
There are two or three 180F-drive boards underneath the instrument. Each has a spe-
cific role: one board controls the drives, slits, shutter, and automated accessories in the
excitation monochromator. Another board controls the same components of the emis-
sion monochromator. The third board is an optional sample-compartment-accessory
control board for options such as a sample changer. (See photograph on next page)
Do not detach
screens’ screws
CTI card
CPU board
CTI Card
CPU board
Older version
Newer version
Summary of Contents for FluoroMax-3
Page 5: ...FluoroMax 3 v 3 1 3 Mar 2006 v 15 Declaration of Conformity 15 1 16 Index 16 1...
Page 6: ...FluoroMax 3 v 3 1 3 Mar 2006 vi...
Page 18: ...FluoroMax 3 v 3 1 3 Mar 2006 Introduction 0 12...
Page 36: ...FluoroMax 3 v 3 22 Feb 2005 System Description 2 8...
Page 96: ...FluoroMax 3 v 3 1 3 Mar 2006 Maintenance 6 12...
Page 156: ...FluoroMax 3 v 3 1 6 Mar 2006 Automated Polarizers 10 22...
Page 208: ...FluoroMax 3 v 3 1 6 Mar 2006 Bibliography 14 4...
Page 210: ...FluoroMax 3 v 3 1 6 Mar 2006 Declaration of Conformity 15 2...