FluoroMax-3 v. 3.1 (7 Mar 2006)
Producing Correction Factors
8-8
Note:
Doing so replaces the
xcorrect
file
that was shipped with the system.
b
Click the
Detectors
icon
.
c
Enable both
S1
and
R1
detectors.
The codes
S1
and
R1
appear in the
Formulas
table.
d
Remove
both
S1
and
R1
from the
Formulas
table.
e
In the
Signal
column, click
S1
. In the
Operations
column, click the
division sign,
/
. In the
Signal
column, click
R1
.
In the
Formula
box,
S1/R1
appears.
f
Click
Add >>
to add this formula to the
Formulas
table.
g
Enter an
Integration Time
of 1 second.
5
Click Run to execute the scan.
6
Save the
file as
xcorrect
.
7
Normalize the data in the same way as in the
previous section.