FluoroMax-3 v. 3.1 (3 Mar 2006)
System Operation
3-5
5
Use the default parameters or adjust them.
Default monochromator parameters for the xenon-lamp scan
Monochromator
(1200
grooves/mm)
Initial wave-
length
Final wave-
length
Increment Slits
(bandpass)
Excitation
200 nm
600 nm
1 nm
1 nm
Emission
350 nm
--
--
1 nm
Default detector parameters for the xenon-lamp scan
Detector (Signal)
Integration time
Units
Signal (S1)
100 ms
CPS
Reference (R1)
100 ms
mA
6
Click Run.
The
Intermediate Display
opens. The xenon-lamp scan runs: