FluoroMax-3 v. 3.1 (3 Mar 2006)
System Operation
3-4
Excitation calibration check
This calibration check verifies the wavelength calibration of your excitation mono-
chromator, using the reference photodiode located before the sample compartment. It is
an excitation scan of the xenon lamp’s output, and should be the first check performed.
1
Close the lid of the sample compartment.
2
On the main
FluorEssence
toolbar, select the Ex-
periment Menu button:
The
Fluoresence Main
Experiment Menu
appears:
3
Choose Spectra.
The
Experiment Type
window appears:
4
Choose Excitation.
The xenon-lamp scan experiment
automatically loads.