Notes on excitation and emission calibration
•
Two experiments,
lamp.exp
and
water.exp
, have been defined and saved.
They can be run, after the system is switched on each day, to check the calibration
and performance of the FluoroMax
®
-3.
•
HORIBA Jobin Yvon Inc. recommends monitoring the number of hours of xenon-
lamp use, via the hour meter.
•
Additionally, you may want to record the water-Raman intensity daily or weekly.
•
The lamp is rated for 1200–1500 h, but if the Raman intensity starts to drop, you
may wish to change the lamp sooner.