FluoroMax-3 v. 3.1 (3 Mar 2006)
System Operation
3-12
6
Use the default parameters or adjust them.
Monochromator parameters for the water-Raman scan:
Monochromator
(1200 grooves/mm)
Initial wave-
length
Final wave-
length
Incre-
ment
Slits (band-
pass)
Excitation
350 nm
--
--
5 nm
Emission
365 nm
450 nm
1 nm
5 nm
Detector parameters for the water-Raman scan:
Detector (Signal)
Integration time
Units
Signal (S1)
100 ms
CPS
Reference (R1)
100 ms
mA
7
Click Run.
The
Intermediate Display
opens. The water-Raman scan runs.
A new FluoroMax
®
should display a water-Raman peak intensity of at least
300 000 counts s
–1
.