40
6.2
Running Tests
6.2.1
Accessing the Tests Menu
List of Tests
You must systematically run tests after calibrating sensors.
At the end of a full test cycle, you can obtain certificates. See section 6.3.2, page 66.
Calibration
Associated Tests
Force sensor
Occlusion alarm
Syringe barrel clasp
Syringe barrel clasp
Displacement
Displacement
1.
Access the
Options
menu. See section 3.3, page 29.
2.
Press the arrow keys to select
Maintenance
.
3.
Press
enter
.
4.
Enter the Maintenance access code.
5.
Press
OK
.
6.
Press the arrow keys to select
Tests
.
7.
Press
enter
.
The
Tests
screen is displayed.
Tests
Tests
Test 1
Identification. See section 6.2.2, page 41.
Test 11
Ageing test. See section 6.2.12, page 52.
Test 2
Maintenance options. See section 6.2.3, page 42.
Test 12
Battery life. See section 6.2.13, page 53.
Test 3
Maintenance messages.
See section 6.2.4, page 42.
Test 13
Battery test. See section 6.2.14, page 53.
Test 4
Events. See section 6.2.5, page 42.
Test 20
Pump code. See section 6.2.15, page 55.
Test 5
Force. See section 6.2.6, page 43.
Test 21
Power information. See section 6.2.16, page 55.
Test 6
Displacement. See section 6.2.7, page 45.
Test 22
LCD voltage. See section 6.2.17, page 57.
Test 7
Syringe barrel clasp. See section 6.2.8, page 47.
Test 23
Temperature. See section 6.2.18, page 57.
Test 8
Battery information. See section 6.2.9, page 50.
Test 24
Watchdog. See section 6.2.19, page 58.
Test 9
Indication. See section 6.2.10, page 51.
Test 25
Wi-Fi module information.
See section 6.2.20, page 59.
Test 10
Summary of Contents for Agilia SP
Page 139: ...139 17 Spare Parts Catalog Refer to the spare parts catalog of the relevant pump ...
Page 143: ...143 VTBI Volume to Be Infused W Watts Term Description ...
Page 147: ...Local Contacts for Servicing Write your contacts here ...
Page 148: ...11378 2_tm_agilia_sp_eng ...
Page 305: ...Local Contacts for Servicing Write your contacts here ...