119
16.2
Recommended and Prohibited Agents
16.2.1
Recommended Agents
16.2.2
Prohibited Agents
16.3
Instructions for Cleaning and Disinfecting
We recommend the following cleaning and disinfecting agents:
Recommended Agent
Cleaning
Didecyldimethylammonium chloride
(example: Wip’Anios Excel by
Anios
)
Disinfecting
Didecyldimethylammonium chloride
(example: Wip’Anios Excel by
Anios
)
The following cleaning and disinfecting agents are prohibited:
Trichloroethylene
Abrasive detergents
Undiluted alcohol
These aggressive agents may damage the plastic parts of the pump and
cause it to malfunction.
Follow the instructions provided to ensure effective cleaning and
disinfecting of the equipment.
Use the agents according to the manufacturer's instructions. This
may include wearing personal protective equipment (gloves, lab
coat, glasses, and so on), or diluting the agent according to the
manufacturer's guidelines.
For disinfectants, respect the contact time required for the
antimicrobial agents to act (the time the agent must be left on the
pump for disinfection to be effective).
The following warning is provided to protect staff against electric shock,
and to protect the pump from damage that can cause it to malfunction.
Warning
Only trained staff can clean and disinfect the pump.
Do not place the pump in an autoclave or immerse it in liquid.
Do not spray liquids directly on connectors. Instead, use a
cleaning cloth or disposable wipes.
Summary of Contents for Agilia SP
Page 139: ...139 17 Spare Parts Catalog Refer to the spare parts catalog of the relevant pump ...
Page 143: ...143 VTBI Volume to Be Infused W Watts Term Description ...
Page 147: ...Local Contacts for Servicing Write your contacts here ...
Page 148: ...11378 2_tm_agilia_sp_eng ...
Page 305: ...Local Contacts for Servicing Write your contacts here ...