![Fresenius Kabi Agilia SP Technical Manual Download Page 180](http://html.mh-extra.com/html/fresenius-kabi/agilia-sp/agilia-sp_technical-manual_2331626180.webp)
32
6.2
Using the Pump for the First Time
6.3
Powering on
1.
Make sure the pump is correctly installed at the bedside.
2.
Plug the pump into the AC power supply.
3.
Before starting the pump for the first time, you must charge the
battery for approximately
6 hours
.
Wait until the pump is fully charged.
Do not use the pump during the first charge.
4.
Power on the pump.
5.
Install a syringe into the pump.
Information
The pump can operate using the battery; however, we
recommend that the pump be connected to a power supply as
often as possible during use in order to ensure that the battery
remains charged.
When the pump is connected to the power supply, check that the
power supply indicator
lights up green, and that the power
cord and the wall plug are accessible.
1.
Press
.
An auto-test checks the functionality of the pump.
2.
Immediately after powering on the pump, make sure that all LED
lights blink.
3.
Acknowledge the different screens listed in the table below.