![Fresenius Kabi Agilia SP Technical Manual Download Page 56](http://html.mh-extra.com/html/fresenius-kabi/agilia-sp/agilia-sp_technical-manual_2331626056.webp)
56
1.
Access the
Tests
menu. See section 6.2.1, page 40.
2.
Press the arrow keys to select
Test 21: Power info
.
3.
Press
enter
.
The DC voltage is displayed:
Conversion
: DC voltage conversion value (in mV)
DC voltage
: DC voltage charging value (in V)
4.
Connect and disconnect the pump from the power source, and check that the DC voltage value
changes:
With a pump connected to an AC power supply:
With a pump running on battery:
5.
Press the arrow keys to scroll through the other information screens
:
Conversion
: Backup capacitor conversion value in mV
Backup capacitor
: Backup capacitor voltage in V
Conversion
: Backup capacitor temperature conversion value in mV
Backup temperature
: Backup capacitor temperature in °C
Conversion
: 5 V booster conversion value in mV
5 V booster
: 5 V booster voltage in V
6.
Check that the displayed values are within the limits shown in the table below.
Value
Limits
DC Voltage - Pump connected to an AC power supply
9.5 V to 10.5 V
DC Voltage - Pump running on battery
-0.5 V to 0.5 V
Backup capacitor
2.3 V to 2.7 V
Backup temperature
-20 °C to +110 °C (-4 °F to +230 °F)
5 V booster
4600 mV to 5920 mV
7.
Press
exit
to return to the
Tests
menu.
Summary of Contents for Agilia SP
Page 139: ...139 17 Spare Parts Catalog Refer to the spare parts catalog of the relevant pump ...
Page 143: ...143 VTBI Volume to Be Infused W Watts Term Description ...
Page 147: ...Local Contacts for Servicing Write your contacts here ...
Page 148: ...11378 2_tm_agilia_sp_eng ...
Page 305: ...Local Contacts for Servicing Write your contacts here ...