8.13
Procedure #13: Plunger Cover and Lever Kit
Required Tools and Equipment
1 Torx T-10 screwdriver
1 Torx T-8 screwdriver
1 flat screwdriver
1 anti-static bracelet
Maintenance level
Level 3 specialist technician.
Prerequisites
The battery is disconnected and the boost capacitor discharged.
See section 8.1, page 79.
Dismantling
1.
Turn the pump upside down.
2.
Unscrew and remove the 2 screws [1] that hold the syringe guard.
3.
Remove the syringe guard [2].
4.
Remove the 2 plastic caps [3] using a flat screwdriver.
5.
Unscrew then remove the 2 screws that hold the plunger cover in place.
6.
Remove the plunger cover [4] and lever kit [5].
1
1
2
3
3
5
4