76
28
Keypad
Run test 10 to check the keypad. See section 6.2.11, page 51.
Replace the upper case. See section 8.2, page 81.
Run test 10 to check the keypad.
If the problem cannot be resolved, replace the Display board.
See section 8.3, page 83.
29
Analog/digital converter
Replace the CPU board.See section 8.4, page 85.
Carry out a complete configuration/calibration/control cycle using
Agilia Partner maintenance software.
30
Time keeper
Replace the Power supply board supporting the secondary
microcontroller. See section 8.5, page 88.
Replace the CPU board. See section 8.4, page 85.
Carry out a complete configuration/calibration/control cycle using
Agilia Partner maintenance software.
31
Syringe parameters
Carry out a complete configuration using Agilia Partner
maintenance software.
33
Disengagement optical
switch
Check the functionality of the disengagement switch.
Check the cable connection to CPU board (J9 connector).
See figure 2.8, page 16.
Replace the disengagement flexible circuit.
See section 8.10, page 105.
34
Infusion flow rate
Carry out a complete configuration using Agilia Partner
maintenance software.
35
Motor period
Carry out a complete configuration using Agilia Partner
maintenance software.
36
Motor rotation direction
Carry out a complete configuration using Agilia Partner
maintenance software.
37
OFF key.
Reading status problem or
key pressed too repeatedly
Check that you have not pressed the OFF key too repeatedly.
Replace the Power supply board to check the functionality of the
secondary microcontroller. See section 8.5, page 88.
Replace the Display board. See section 8.3, page 83.
Replace the CPU board to check the functionality of the main
microcontroller.See section 8.4, page 85.
Carry out a complete configuration/calibration/control cycle using
Agilia Partner maintenance software.
38
Locked OFF key
Run test 10 to check the functionality of the keypad, and
especially the OFF key. See section 6.2.11, page 51.
Replace the upper case. See section 8.2, page 81.
Replace the Power supply board to check the secondary
microcontroller functionality. See section 8.5, page 88.
39
Impossible to power off
Replace the Power supply board.See section 8.5, page 88.
40
TCI calculation error
(for Agilia SP TIVA)
Restart the pump.
Program and start an infusion with a TCI profile.
Please refer to the Agilia SP TIVA Instructions for Use.
43
LCD communication failure
Replace the Display board. See section 8.3, page 83.
44
LCD driver failure
Replace the Display board. See section 8.3, page 83.
45
Defective LCD voltage
Replace the Display board.See section 8.3, page 83.
46
Backup capacitor
temperature too high
Contact the
Fresenius Kabi
Technical Service.
47
Backup capacitor
temperature out of range
Contact the
Fresenius Kabi
Technical Service.
48
Pump temperature out of
range
Place the pump in normal operating temperature.
Please refer to the Instructions for Use of the relevant pump.
49
Backup capacitor and safety
5 V booster charging circuit
defective
Replace the Power supply board. See section 8.5, page 88.
50
Power supply board voltage
out of range
Replace the Power supply board. See section 8.5, page 88.
Error Code
Description
Recommended Actions
Summary of Contents for Agilia SP
Page 139: ...139 17 Spare Parts Catalog Refer to the spare parts catalog of the relevant pump ...
Page 143: ...143 VTBI Volume to Be Infused W Watts Term Description ...
Page 147: ...Local Contacts for Servicing Write your contacts here ...
Page 148: ...11378 2_tm_agilia_sp_eng ...
Page 305: ...Local Contacts for Servicing Write your contacts here ...