![Fresenius Kabi Agilia SP Technical Manual Download Page 200](http://html.mh-extra.com/html/fresenius-kabi/agilia-sp/agilia-sp_technical-manual_2331626200.webp)
52
7.9
Functions During Infusion
7.9.1
Stop
7.9.2
Rate Titration
To stop the infusion, press
.
After 2 minutes, an alarm is generated as a reminder that the infusion is
stopped.
To restart the infusion, first confirm or modify the programming settings,
then start the infusion. See section 7.6, page 41.
You can adjust the infusion rate (flow rate or dose) during the infusion.
Depending on your pump configuration, stopping the infusion may be
required before modifying the infusion rate.
1.
If required, stop the infusion, see section 7.9.1, page 52.
2.
Press the arrow keys to modify the flow rate or dose.
3.
Press
OK
to confirm.
Flow rate
Dose
Flow rate
Dose