100
10.
Disconnect the cable that connects the CPU board to displacement linear potentiometer at
[J2].
11.
Push the disengagement lever [8] down and move the plunger driver to its maximum.
12.
Remove the 2 plastic caps [9] using a flat screwdriver.
13.
Unscrew and remove the 2 screws.
14.
Unclip and remove the syringe flange cradle [10].
15.
Unscrew and remove the 2 screws [11] that hold the base.
16.
Remove the base [12].
While removing the base [12], pay attention to the position of angle bracket clip kit [13]
between the base and the aluminum profile. When re-assembling the base, make sure to
position this part as it was originally.
J2
9
8
9
12
9
11
11
13
13
12