82
8.16
View Pressure History
Symbol
Procedure
Viewing pressure history
This function allows the user to check the current infusion’s history
information in order to verify changes in pressure.
You can view pressure history as follows:
1.
Press
.
2.
Press the arrow keys to select
.
3.
Press
enter
.
The following information is displayed:
- An event marker (cursor)
- The event details (time and pressure limit)
- The pressure limit (dotted line)
- The measured pressure (solid line)
4.
Press the
and
buttons to browse the events.
5.
Press
to view information about the selected event.
Information
The history is not refreshed while the history screen is displayed.
To refresh the history data, exit and select the history again.
Pressure history is not stored after powering off.