![Fresenius Kabi Agilia SP Technical Manual Download Page 103](http://html.mh-extra.com/html/fresenius-kabi/agilia-sp/agilia-sp_technical-manual_2331626103.webp)
103
7.
Disconnect the cable that connects the motor to the CPU board at [J7].
8.
Unscrew and remove the screw [6] that holds the motor ground wire [9] to the flange.
9.
Unscrew and remove the 4 screws [7] that hold the motor [8] to the mechanical framework.
10.
Remove the motor [8].
J7
8
9
7
7
7
7
6
8
Summary of Contents for Agilia SP