126
18.6
Sound Levels
18.6.1
Operating Pump Sound Levels (without alarms)
Note: These values are provided for information purposes only.
18.6.2
Alarms Sound Levels
18.7
Compliance
* After a defibrillation, the pump recovery time is around 2 seconds.
** The functional earth is directly connected to the power supply cord. It reduces residual current that
may disturb ECG or EEG devices.
Flow Rate (mL/h)
Sound Level (dBA)
0
21
1
23
20
27
100
30
400
49
1200
32
Alarm Priority
Sound Level (dBA)
min
Max
High-priority
55
63
Medium-priority
50
57
Low-priority
45
52
ElectroMedical
Equipment Safety
Compliant with the following
standards:
IEC 60601-1
IEC 60601-1-8
IP22
Index of protection against
ingress of water or particulate
matter
EMC
(ElectroMagnetic
Compatibility)
Compliant with the following
standard:
IEC 60601-1-2
Protection against leakage
current: Defibrillation-proof
type CF applied part*
Particular
Standards
IEC 60601-2-24
ANSI/AAMI ID26
Protection against electric
shocks: class II
Functional earth**
Summary of Contents for Agilia SP
Page 139: ...139 17 Spare Parts Catalog Refer to the spare parts catalog of the relevant pump ...
Page 143: ...143 VTBI Volume to Be Infused W Watts Term Description ...
Page 147: ...Local Contacts for Servicing Write your contacts here ...
Page 148: ...11378 2_tm_agilia_sp_eng ...
Page 305: ...Local Contacts for Servicing Write your contacts here ...