59
6.2.20
Test 25: Wi-Fi Module Information
Test Description
This test displays for information the Wi-Fi parameters saved on the pump. Run this test to
check that all parameters are correct.
If they are not, upload the Wi-Fi configuration to the pump using Agilia Partner maintenance
software.
Prerequisites
The pump must be equipped with Wi-Fi.
Information
When powering on a pump, it may take a few minutes for the Wi-Fi module information to refresh.
If some data are not displayed when performing this test, wait a few minutes for the pump to refresh Wi-Fi
module information (without powering off the pump), then run the test again.
If after a prolonged period where no information is displayed, contact your Biomed or IT service
department.
1.
Access the
Tests
menu. See section 6.2.1, page 40.
2.
Press the arrow keys to select
Test 25: Wi-Fi module information
.
3.
Press
enter
.
The following data is displayed for information:
Network SSID (Service Set IDentifier)
MAC address
IP address
Subnet mask address
Gateway address
RSSI: value in dBm
SNR: value in dBm
Data rate: value in Mbps.
Date and time of last connection to server
Version of Wi-Fi module application firmware
Version of Wi-Fi module Bootloader firmware
Version of Wi-Fi module Radio firmware
4.
Press
exit
to return to the
Tests
screen.
Summary of Contents for Agilia SP
Page 139: ...139 17 Spare Parts Catalog Refer to the spare parts catalog of the relevant pump ...
Page 143: ...143 VTBI Volume to Be Infused W Watts Term Description ...
Page 147: ...Local Contacts for Servicing Write your contacts here ...
Page 148: ...11378 2_tm_agilia_sp_eng ...
Page 305: ...Local Contacts for Servicing Write your contacts here ...