![Fresenius Kabi Agilia SP Technical Manual Download Page 133](http://html.mh-extra.com/html/fresenius-kabi/agilia-sp/agilia-sp_technical-manual_2331626133.webp)
133
12.4
Communication Port
12.5
Infrared Communication
12.6
Agilia USB Cable
12.7
Compliance
* After a defibrillation, the pump recovery time is around 2 seconds.
** The functional earth is directly connected to the power suply cord. It reduces residual current that may disturb ECG or EEG devices.
The connector located at the back of the device allows data communication with a PC.
Serial Cable
TTL output
Power Input
10 V / 15 W to power supply the product
Power Output
5 VDC / 150 mA to power Agilia USB cable
The pump is equipped with an infrared cell located at the back of the device.
Mode
Wireless optical communication using infrared light
Compatibility
Asynchronous Serial Infrared (SIR) physical layer irPHY 1.0, baseband no carrier
Transport
Protocol
Proprietary
Speed
115.2 kb/s max
Wavelength
880 nm to 900 nm infrared band with 45 nm spectral bandwidth
Eye Safety
Class 0 of IEC 62471
Function
Allows communication between a pump and a computer on the USB port
Connector
8-pin proprietary connector / Male A USB connector
Input / Output
power (pump)
250 mW
Input / Output
Power (PC)
100 mW
Compatibility
USB 2.0 compliant.
Insulation
1500 V, compliant with UL standard
Speed
19.2 kb/s max
Length
85 in (2.15 m)
Maximum Length
110 in (2.8 m)
ElectroMedical
Equipment Safety
Compliant with the following standards:
IEC 60601-1
IEC 60601-1-8
IP22
Index of protection against solid foreign objects
(> 12.5 mm) and driping liquids
EMC
(ElectroMagnetic
Compatibility)
Compliant with the following standards:
IEC 60601-1-2
Protection against leakage current: Defibrillation-proof
type CF applied part*
Particular
Standards
IEC 60601-2-24
ANSI/AAMI ID26
Protection against electric shocks: class II
Functional earth**
Summary of Contents for Agilia SP
Page 139: ...139 17 Spare Parts Catalog Refer to the spare parts catalog of the relevant pump ...
Page 143: ...143 VTBI Volume to Be Infused W Watts Term Description ...
Page 147: ...Local Contacts for Servicing Write your contacts here ...
Page 148: ...11378 2_tm_agilia_sp_eng ...
Page 305: ...Local Contacts for Servicing Write your contacts here ...