13
1.8
Use Environment
Agilia SP MC is intended for use in healthcare facilities, under the
supervision of trained healthcare personnel.
The pump must be used in the following operational conditions to ensure
proper performance:
Operating temperature range:
5 °C to 40 °C
Operating pressure range:
700 hPa (525 mmHg / 10.15 PSI) to 1060 hPa (795 mmHg /
15.37 PSI)
Operating humidity range:
20 % to 90 % with no condensation
Altitude:
Up to 3000 m above sea-level
Warning
The functionality of the pump can be affected by pressure variations,
mechanical shocks, heat ignition sources, and so on.
Information
The pump can be used in ambulances exclusively with the Agilia
Holder Ambulance accessory. Due to use in road ambulances,
performances of the device can be modified. For more
information, refer to Agilia Holder Ambulance IFU.
For more information on using the device in specific conditions,
contact your
Fresenius Kabi
representative.