141
23.4.2
Table 2 - Guidance and Manufacturer's Declaration -
Electromagnetic Immunity
Warning
Agilia pump and its accessories are intended to be used in the
electromagnetic environments specified below.
The customer or the user of Agilia pump should ensure that it is
used in such environments.
Immunity Test
IEC 60601-1-2
- - - - - - - - - - - - - -
Compliance
Level
Obtained by
the Device
Electromagnetic Environment -
Guidance
IEC 60601-2-24
and ANSI/AAMI
ID26
Test Level
Electrostatic
Discharge (ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
- - - - - - - - - - - - - -
± 8 kV contact
± 15 kV air
Floor coverings made from wood, tiles and
concrete, with relative humidity level at
least 30 %, make it possible to guarantee
the necessary level of conformity. If it is not
possible to guarantee this environment,
additional precautions must be taken, such
as: use of anti-static equipment,
preliminary user discharge and the
wearing of antistatic clothing.
± 8 kV contact
± 15 kV air
Electrical fast
Transient / burst
IEC 61000-4-4
± 2 kV for power
supply lines
± 1 kV for input
output lines
± 2 kV for power
supply lines
± 1 kV for input
output lines
AC power quality should be that of a
typical commercial or healthcare facility
environment.
Surge
IEC 61000-4-5
± 1 kV
differential mode
± 2 kV
common mode
± 1 kV
differential mode
± 2 kV
common mode
AC power quality should be that of a
typical commercial or healthcare facility
environment.
Voltage dips,
short
interruptions and
voltage
variations on
power supply
input lines
IEC 61000-4-11
< 5% Ut
(> 95% dip in Ut)
for 0.5 cycles
< 5% Ut
(> 95% dip in Ut)
for 0.5 cycles
AC power quality should be that of a
typical commercial or healthcare facility
environment.
For short and long interruptions (< than
battery life) of AC power, the internal
battery provides continuity of service.
40% Ut
(60% dip in Ut)
for 5 cycles
40% Ut
(60% dip in Ut)
for 5 cycles
70% Ut
(30% dip in Ut)
for 25 cycles
70% Ut
(30% dip in Ut)
for 25 cycles
< 5% Ut
(> 95% dip in Ut)
for 5 s
< 5% Ut
(> 95% dip in Ut)
for 5 s
Summary of Contents for Agilia SP
Page 139: ...139 17 Spare Parts Catalog Refer to the spare parts catalog of the relevant pump ...
Page 143: ...143 VTBI Volume to Be Infused W Watts Term Description ...
Page 147: ...Local Contacts for Servicing Write your contacts here ...
Page 148: ...11378 2_tm_agilia_sp_eng ...
Page 305: ...Local Contacts for Servicing Write your contacts here ...