Figure 6.3:
Installation of V1 diameter verification gauge
Figure 6.4:
Installation of V2 diameter verification gauge
4.
Press the arrow keys to scroll through the information screens
.
The following data is displayed:
Conversion
: Sensor value in mV. The displayed voltage value depends on
the syringe diameter installed
Diameter
: Syringe diameter in mm. Value measured by the sensor.
Low (11 mm)
: Sensor value in mV when the C1 calibration tool
(diameter = 11 mm) is installed. Value saved during last calibration
Middle (22.9 mm)
: Sensor value in mV when the C2 calibration tool
(diameter = 22.9 mm) is installed. Value saved during last calibration
High (33 mm)
: Sensor value in mV when the C3 calibration tool
(diameter = 33 mm) is installed. Value saved during last calibration
Date
: Date of last calibration
Number
: Last calibration number
5.
Press the arrow keys to display the
Conversion/Diameter
screen.
6.
Open the syringe barrel clasp, push the disengagement lever down, and move the plunger driver to
the right.
7.
Place the V1 diameter verification gauge in the syringe cradle, with the flanges correctly inserted in
the provided slot.
8.
Secure the V1 diameter verification gauge with the syringe barrel clasp, and press
OK
on the pump.
9.
Check the value displayed for
Diameter
. The value must be equal to 10.5 mm ± 1.5 mm
(0.35 in to 0.47 in).
10.
Open the syringe barrel clasp, and remove the V1 diameter verification gauge.
11.
Place the V2 diameter verification gauge in the syringe cradle, with the flanges correctly inserted in
the provided slot.
12.
Secure the V2 diameter verification gauge with the syringe barrel clasp, and press
OK
on the pump.
13.
Check the value displayed for
Diameter
. The value must be equal to 15.5 mm ± 1.5 mm
(0.55 in to 0.70 in).
14.
Open the syringe barrel clasp, and remove the V2 diameter verification gauge.
15.
Place the V4 diameter verification gauge in the syringe cradle, with the flanges correctly inserted in
the provided slot.
16.
Secure the V4 diameter verification gauge with the syringe barrel clasp, and press
OK
on the pump.
Summary of Contents for Agilia SP
Page 139: ...139 17 Spare Parts Catalog Refer to the spare parts catalog of the relevant pump ...
Page 143: ...143 VTBI Volume to Be Infused W Watts Term Description ...
Page 147: ...Local Contacts for Servicing Write your contacts here ...
Page 148: ...11378 2_tm_agilia_sp_eng ...
Page 305: ...Local Contacts for Servicing Write your contacts here ...