35
6.5
Pump Height
Precautions for pump position
If using multiple syringe pumps and it is not clinically feasible to have
all pumps level with the distal tip of the catheter (or the site of fluid
delivery), place the high risk or life-sustaining medications as close to
level with the distal tip of the catheter as possible. When infusing multiple
high risk or life-sustaining medications, consider placing the ones
infusing at the lowest rates as close to the level with the distal tip of the
catheter as possible.
Minimize the height difference between the pump and the patient and
avoid changes in the height of the pump (e.g., during transport of
critically ill patients) to prevent unintended fluctuations in the flow rate.
Warning
Ideally, the syringe pump should be level with the distal tip of the
catheter (e.g., the site of fluid delivery; if accessing a central line the
syringe pump should be at the level of the patient’s heart). If the pump
height is raised relative to the distal tip of the catheter (e.g., during
patient transport), the increase in height of the syringe pump can result
in a temporary increase in fluid delivery or bolus until the flow rate
stabilizes. Alternatively, if the pump is lowered relative to the distal tip of
the catheter, the decrease in height of the syringe pump may result in a
decrease in delivery or under-infusion until the flow rate stabilizes.
Summary of Contents for Agilia SP
Page 139: ...139 17 Spare Parts Catalog Refer to the spare parts catalog of the relevant pump ...
Page 143: ...143 VTBI Volume to Be Infused W Watts Term Description ...
Page 147: ...Local Contacts for Servicing Write your contacts here ...
Page 148: ...11378 2_tm_agilia_sp_eng ...
Page 305: ...Local Contacts for Servicing Write your contacts here ...