58
6.2.19
Test 24: Watchdog
Test Description
This test checks the status of the pump’s microprocessor (free or busy) every 2 seconds.
The pump’s microprocessor is not too busy to carry out infusions correctly.
If the test fails, it means that the pump’s microprocessor is too busy.
An infusion is likely not to be carried out correctly.
Required Tools and Equipment
No specific tools or equipment are required
Prerequisites
The pump must be used on battery with the AC power cord disconnected.
1.
Access the
Tests
menu. See section 6.2.1, page 40.
2.
Press the arrow keys to select
Test 24: Watchdog
.
3.
Press
enter
.
If the pump is connected to the AC power supply:
Disconnect the AC power cable.
The
OK
button becomes available.
Press
OK
.
4.
Press
OK
to confirm.
The pump starts checking the fault mode.
5.
Check that all red LEDs light up and that the buzzer sounds between 1.0 seconds and 2.25 seconds.
If the test fails, a message "Watchdog not functional return to After Sales Service" is displayed.
6.
Press key
for 5 seconds to stop the alarm and return to
Tests
screen.
Information
The Watchdog test must be systematically carried out after each maintenance procedure.