![Fresenius Kabi Agilia SP Technical Manual Download Page 259](http://html.mh-extra.com/html/fresenius-kabi/agilia-sp/agilia-sp_technical-manual_2331626259.webp)
111
15.2
Flow Rate
* Direct bolus default value = Upper flow rate value for each syringe size.
** KVO defaut value = 1 mL/h.
15.3
Volume To Be Infused (VTBI)
Applicable for all syringe sizes (50/60 mL, 30 mL, 20 mL, 10 mL, 5 mL)
Format
Syringe
Minimum
Increment
50 mL
/
60 mL
30 mL 20 mL 10 mL
5 mL
Infusion Rate
mL/h
0.1
1200
0.1
600
0.1
600
0.1
350
0.1
250
0.01
0.1
1
(0.10
9.99)
(10.0
99.9)
(100
1200)
Direct Bolus*
mL/h
50
1200
50
600
50
600
50
350
50
250
50
Priming
mL/h
1200
600
600
350
250
N/A
Programmed Bolus
mL/h
0.1
1200
0.1
600
0.1
600
0.1
350
0.1
250
0.01
0.1
1
(0.10
9.99)
(10.0
99.9)
(100
1200)
KVO**
mL/h
0.1
5
0.1
5
0.1
5
0.1
5
0.1
5
0.1
Loading Dose
mL/h
0.1
1200
0.1
600
0.1
600
0.1
350
0.1
250
0.01
0.1
1
(0.10
9.99)
(10.0
99.9)
(100
1200)
Format
Range of Settings
Default
Value
Minimum
Increment
Volume Limit
mL
0.1
999
0.1
0.1
1
(0.1
99.9)
(100
999)
Volume/Time
mL
0.1
99.9
0.1
0.1
Direct Bolus
mL
0.1
50
N/A
0.1
Programmed Bolus
mL
0.1
99.9
0.1
0.1
Summary of Contents for Agilia SP
Page 139: ...139 17 Spare Parts Catalog Refer to the spare parts catalog of the relevant pump ...
Page 143: ...143 VTBI Volume to Be Infused W Watts Term Description ...
Page 147: ...Local Contacts for Servicing Write your contacts here ...
Page 148: ...11378 2_tm_agilia_sp_eng ...
Page 305: ...Local Contacts for Servicing Write your contacts here ...