![Fresenius Kabi Agilia SP Technical Manual Download Page 246](http://html.mh-extra.com/html/fresenius-kabi/agilia-sp/agilia-sp_technical-manual_2331626246.webp)
98
12.4.2
Volume Limit Alarms
12.4.3
Volume/Time & Dose/Time Alarms
12.4.4
Infusion Alarms
Message
Priority
Stops
Infusion?
Problem / Resolution
End of volume
limit !!!
High (!!!)
Yes
The volume limit is reached.
Near end of volume
limit !!
Medium (!!)
No
The time remaining is less than the defined
time duration (adjustable between 1 and
30 minutes), and the remaining VTBI until the
volume limit has dropped to less than 10% of
the syringe capacity.
End of volume
limit !
Low (!)
No
The volume limit is reached and the end of
infusion setting is set as "KVO" or
"continuous".
Message
Priority
Stops
Infusion?
Problem / Resolution
End of
volume/time !!!
High (!!!)
Yes
The VTBI is completed.
End of dose/time !!!
Near end of
volume/time !!
Medium (!!)
No
The time remaining is less than the defined
time duration (adjustable between 1 and
30 minutes), and the remaining VTBI has
dropped to less than 10 % of the syringe
capacity.
Near end of
dose/time !!
End of volume/time !
Low (!)
No
The VTBI is completed and the end of infusion
setting is set as "KVO" or "continuous".
End of dose/time !
Message
Priority
Stops
Infusion?
Problem / Resolution
End of infusion !!!
High (!!!)
Yes
The infusion is completed (simple rate).
Near end of
infusion !!
Medium (!!)
No
The time remaining is less than the defined
time duration (adjustable between 1 and
30 minutes), and the remaining volume of fluid
in the syringe has dropped to less than 10 %
of the syringe capacity.
Check settings !!
Medium (!!)
No
The flow rate (or dose) has been modified
using the keys, but has not been confirmed.
Check the flow rate (or dose) and press
OK
to confirm.
Waiting settings !!
Medium (!!)
No
A value must be entered.
Enter a value and press
OK
to confirm.
Summary of Contents for Agilia SP
Page 139: ...139 17 Spare Parts Catalog Refer to the spare parts catalog of the relevant pump ...
Page 143: ...143 VTBI Volume to Be Infused W Watts Term Description ...
Page 147: ...Local Contacts for Servicing Write your contacts here ...
Page 148: ...11378 2_tm_agilia_sp_eng ...
Page 305: ...Local Contacts for Servicing Write your contacts here ...