90
Re-assembling
1.
Place the black insulator part [7] on the AC supply board.
2.
Insert the power supply board in slot [6].
3.
Locate securing clip [5] then press the power supply board down, locking the board into
position with the securing clip.
4.
Perform the dismantling procedure in reverse order to re-assemble the unit.
Pay attention to the direction of the flat cable connections when assembling.
5.
Perform the regular servicing tests. See Quality Control Procedure, page 63.