![Fresenius Kabi Agilia SP Technical Manual Download Page 50](http://html.mh-extra.com/html/fresenius-kabi/agilia-sp/agilia-sp_technical-manual_2331626050.webp)
50
6.2.9
Test 8: Battery Info
Test Description
This test provides information about battery usage time, voltage output, current output, and
temperature.
Run this test to check that the battery data is correct.
The battery capacity is acceptable.
If this test fails:
Connect the pump to an AC power supply to charge the battery.
Run the test again. If the test fails, replace the battery.
See section 8.1, page 79.
Required Tools and Equipment
No specific tools or equipment are required
1.
Access the
Tests
menu. See section 6.2.1, page 40.
2.
Press the arrow keys to select
Test 8: Battery info
.
3.
Press
enter
.
4.
Press the arrow keys to scroll through the information screens
.
The following data is displayed:
Duration
: Battery operating time in hours, days and months
Cycles
: Number of charge/discharge cycles, equivalent to Duration (in
hours) divided by 7
Since
: Date of the first use on battery
Conversion
and
Voltage
: Voltage value at the battery terminals in V, and
conversion into mV
Conversion
and
Temperature
: Battery temperature in °C, and conversion
into LSB
Conversion
and
Current
: Battery charge/discharge current in mA, and
conversion into LSB
5.
Check that the displayed values are within the limits shown in the table below.
Value
Limits
Voltage
5.0 V to 8.7 V
Current - Connected to an AC power supply, battery being charged
(battery charge status indicator is flashing)
-135 mA to -165 mA
Current - Connected to an AC power supply, battery fully charged
(battery charge status indicator is lit permanently)
-50 mA to +50 mA
Current - Pump running on battery
(battery charge status indicator is off)
-50 mA to -200 mA
Temperature
+2 °C to +48 °C (+35.6 °F to +118.4 °F) for normal
room temperatures
6.
Press
exit
to return to the
Tests
menu.
Summary of Contents for Agilia SP
Page 139: ...139 17 Spare Parts Catalog Refer to the spare parts catalog of the relevant pump ...
Page 143: ...143 VTBI Volume to Be Infused W Watts Term Description ...
Page 147: ...Local Contacts for Servicing Write your contacts here ...
Page 148: ...11378 2_tm_agilia_sp_eng ...
Page 305: ...Local Contacts for Servicing Write your contacts here ...