![Fresenius Kabi Agilia SP Technical Manual Download Page 272](http://html.mh-extra.com/html/fresenius-kabi/agilia-sp/agilia-sp_technical-manual_2331626272.webp)
124
18
Technical Characteristics
18.1
Power Supply
18.2
Battery
= Wi-Fi enabled
= Wi-Fi disabled or not used
18.3
Power Consumption
It is mandatory to use an Agilia power cord compliant with USA
standards and with the IEC 60227 standard.
The power cord conductor must have a cross section of at least
0.75 mm
2
.
AC Power
Power supply
100 V - 240 V ~ / 50 / 60 Hz with functional
earth
Maximum consumption
10-15 VA
Protective fuse
1 X T1.6AH 250V accessible in the battery
compartment
Disconnect the battery before opening the device. Avoid short circuits
and extreme temperatures.
If the device is not used for more than 3 months, the date is erased (all
other settings are stored permanently). When you power on the pump,
you must set the date again.
Characteristics
7.2 V 2.2 Ah - Li-ion Smart battery
Weight
Approximately 100 g
Battery Life
Flow Rate
Wi-Fi
Battery Life
5 mL/h
5 mL/h
> 9 h
> 13 h
Battery Recharge
Pump OFF: < 6 h / Pump ON: < 20 h
The pump typically consumes about 3.5 W in standard operating
conditions.
Summary of Contents for Agilia SP
Page 139: ...139 17 Spare Parts Catalog Refer to the spare parts catalog of the relevant pump ...
Page 143: ...143 VTBI Volume to Be Infused W Watts Term Description ...
Page 147: ...Local Contacts for Servicing Write your contacts here ...
Page 148: ...11378 2_tm_agilia_sp_eng ...
Page 305: ...Local Contacts for Servicing Write your contacts here ...