48
Overriding a Soft Limit
1.
If you reach a soft limit when programming an infusion, the pump
displays a message at the top of the screen:
Upper soft max
= the upper soft limit is exceeded
Lower soft min
= the lower soft limit is exceeded
2.
If the displayed settings correspond to the intended flow rate or
dose, press
OK
.
3.
Carefully review the program settings.
The original infusion settings continue until you confirm the new
settings.
4.
Press
OK
or
start
to confirm the soft limit override.
During infusion, the upper or lower soft limit message will alternate with
the drug name and concentration at the top of the screen.
Flow rate
Dose
Flow rate
Dose
Flow rate
Dose