36
Par 18:
Day/Night mode
Screen brightness low
Brightens or darkens the screen at
next night mode startup.
Enabled / Disabled
Infusion indicators low
Brightens or darkens the infusion
indicators at next night mode startup.
Enabled / Disabled
Key press sound off
Turns on or off the key press sound
at next night mode startup.
Enabled / Disabled
Manual mode
Displays or hides "Day/Night mode"
menu item.
Enabled / Disabled
Auto mode
Enables/disables the ability to
automatically activate night mode.
Enabled / Disabled
Day/Night mode
If
Auto mode
is enabled, press
OK
to set a daily period for the automatic
activation of night mode.
00h00
23h59
00:00
23:59
Par 19:
Other functions
Direct bolus
Activates or deactivates the
Direct
bolus
function.
Enabled / Disabled
Loading dose or
Induction dose
(Agilia SP TIVA)
Activates or deactivates the
Loading
dose
or
Induction dose
function.
Enabled / Disabled
Programmed bolus
Activates or deactivates the
Programmed bolus
function.
Enabled / Disabled
Par 20:
Authorized units
Concentration units
Sets the available concentration
units: nanog, µg, mg, g, mmol, munit,
unit, cal, kcal, mEq, /mL, /XmL
Enabled / Disabled
Dose unit
Sets the available dose units:
nanog/h, nanog/kg/min, nanog/kg/h,
µg/h, µg/min, µg/kg/min, µg/kg/h,
mg/min, mg/h, mg/24h, mg/kg/min,
mg/kg/h, mg/kg/24h, mg/m
2
/h,
mg/m
2
/24h, g/h, g/kg/min, g/kg/h,
g/kg/24h, mmol/h, mmol/kg/h,
mmol/kg/24h, munit/min,
munit/kg/min, munit/kg/h, unit/min,
unit/h, unit/kg/min, unit/kg/h, kcal/h,
kcal/24h, kcal/kg/h, mEq/min,
mEq/h, mEq/kg/min, mEq/kg/h, /mL,
--mL, mL/kg/h, mL/kg/24h
Enabled / Disabled
Par 22:
Programming modes
Dose
Sets the available modes for the
profile.
Enabled / Disabled
Par 23:
Patient parameters
Age (years)
Sets the patient age.
From 1 year to 150 years
Min weight (kg)
Sets the lower programmable limit of
patient weight.
0.25 kg
350 kg
Max weight (kg)
Sets the upper programmable limit of
patient weight.
"Min weight (kg)"
350 kg
Default weight (kg)
Sets the patient default weight.
"Min weight (kg)"
"Max weight (kg)"
Min height (cm)
Sets the lower patient height.
20 cm
250 cm
Max height (cm)
Sets the upper patient height.
"Min height (cm)"
250
cm
Default height (cm)
Sets the default patient height.
"Min height (cm)"
"Max height (cm)"
Min surface (m²)
Sets the lower programmable limit of
patient BSA.
0.05 m²
4.5 m²
Max surface (m²)
Sets the upper programmable limit of
patient BSA.
"Min surface (m²)"
4.5 m²
Default surface (m²)
Sets the default patient BSA.
"Min surface (m²)"
"Max surface (m²)"
Function
Setting
Description
Range of Settings
Summary of Contents for Agilia SP
Page 139: ...139 17 Spare Parts Catalog Refer to the spare parts catalog of the relevant pump ...
Page 143: ...143 VTBI Volume to Be Infused W Watts Term Description ...
Page 147: ...Local Contacts for Servicing Write your contacts here ...
Page 148: ...11378 2_tm_agilia_sp_eng ...
Page 305: ...Local Contacts for Servicing Write your contacts here ...