15.9
Accuracy
15.9.1
Flow Rate Accuracy
15.9.2
Effects of Pressure Variations on Accuracy
15.9.3
Occlusion Alarm Accuracy
* Test condition:
Temperature:
20 °C,
Extension set length:
150 cm
**
Syringe
: BD Precise
Warning
Accuracy (flow rate, time, volume infused, pressure) can be influenced
by syringe model, syringe configuration, extension set configuration,
fluid viscosity, and fluid temperature.
Note: All tests below are in accordance with the IEC 60601-2-24
standard and ANSI/AAMI ID26.
Values are representative of syringes used during internal tests and are
provided as indicators only.
Accuracy
Flow Rate
± 3%
Back Pressure
Accuracy
(From Mean Values)
+ 39.9 kPa
+ 13.33 kPa
- 13.33 kPa
~ - 3%
~ - 1.5%
~ + 1.5%
Accuracy
Occlusion Alarm Response
Time*
Syringe**
Rate
Occlusion Alarm Threshold
50 mmHg
900 mmHg
50 mL
0.1 mL/h
1 mL/h
5 mL/h
< 5 hours
< 30 minutes
< 7 minutes
< 28 hours
< 3 hours
< 30 minutes
20 mL
0.1 mL/h
1 mL/h
5 mL/h
< 4 hours
< 20 minutes
< 5 minutes
< 14 hours
< 75 minutes
< 15 minutes
Summary of Contents for Agilia SP
Page 139: ...139 17 Spare Parts Catalog Refer to the spare parts catalog of the relevant pump ...
Page 143: ...143 VTBI Volume to Be Infused W Watts Term Description ...
Page 147: ...Local Contacts for Servicing Write your contacts here ...
Page 148: ...11378 2_tm_agilia_sp_eng ...
Page 305: ...Local Contacts for Servicing Write your contacts here ...