Wi-Fi
19.1
General Information
The Agilia Connect Infusion System includes an IEEE 802.11 radio-
frequency transmitter incorporated in the Agilia Wi-Fi pumps. It operates
using the following standards and frequencies:
IEEE 802.11a: 5 GHz Frequency Band
IEEE 802.11b: 2.4 GHz Frequency Band
IEEE 802.11g: 2.4 GHz Frequency Band
IEEE 802.11n: 2.4 and 5 GHz Frequency Band
Information
For more information on differentiation between Wi-Fi and non Wi-Fi
pumps, see section 10.3, page 93.
The Wi-Fi module incorporated in the Agilia Wi-Fi pumps is intended to
perform the following, via periodic communication cycles:
Transfer data sets (from Centerium server to pump)
Transfer pump history (from pump to a server)
Communicate general information on the operating status of the
pump
Agilia infusion pumps contain transmitter with the following IDs:
FCC ID: XF6-RSWC301
IC ID: 8407A-RSWC301.
Agilia Wi-Fi pumps must be installed to provide a separation distance of
at least 8 in (20 cm) from all persons and must not be co-located or
operating in conjunction with any other antenna or transmitter.
Warning
Agilia Wi-Fi pumps must be configured by qualified and appropriately
trained staff.
Information
If communication with the wireless network is interrupted, the pump can
be used as intended. For more information, contact your
Fresenius Kabi
sales representative
Summary of Contents for Agilia SP
Page 139: ...139 17 Spare Parts Catalog Refer to the spare parts catalog of the relevant pump ...
Page 143: ...143 VTBI Volume to Be Infused W Watts Term Description ...
Page 147: ...Local Contacts for Servicing Write your contacts here ...
Page 148: ...11378 2_tm_agilia_sp_eng ...
Page 305: ...Local Contacts for Servicing Write your contacts here ...