![Fresenius Kabi Agilia SP Technical Manual Download Page 34](http://html.mh-extra.com/html/fresenius-kabi/agilia-sp/agilia-sp_technical-manual_2331626034.webp)
34
User 4: Pressure
Mode
Selects the mode.
Variable
: One initial pressure
value that can be adjusted during
infusion.
3 levels
: 3 fixed pressure limits
that can be selected during
infusion.
3 levels / Variable
DPS
Activates or deactivates the DPS
(Dynamic Pressure System) mode
while an infusion is in progress.
Yes / No
Unit
Selects the pressure unit.
mmHg / kPa (for
Japanese market only) /
PSI
Maximum pressure value
If
Mode = Variable
, defines the
maximum pressure allowed during
an infusion.
50
900 mmHg
High pressure value
If
Mode = 3 levels
, defines the
maximum pressure allowed during
an infusion.
250
900 mmHg
Medium pressure value
If
Mode = 3 levels
, defines the
medium pressure allowed during an
infusion.
150
600 mmHg
Low pressure value
If
Mode = 3 levels
, defines the low
pressure allowed during an infusion.
50
300 mmHg
Limit stored
If enabled, the last pressure limit
adjustment is automatically stored in
memory for the next startup.
If disabled, the user has to enter
pressure limit for the next startup.
Enabled / Disabled
Limit
If
Limit stored
is disabled, sets the
maximum pressure allowed during
an infusion.
If
Limit stored
is enabled, the limit is
set to ’---’ and cannot be changed.
50 mmHg
"Maximum
pressure value"
Drop threshold
If
DPS = Yes
, sets a pressure drop
threshold. A message is generated if
the pressure drops below this
threshold.
’----’ (No drop pressure
monitoring) or:
100
500 mmHg
Raise threshold
If
DPS = Yes
, defines a pressure rise
threshold. A message is generated if
the pressure rises above this
threshold.
50
400 mmHg
DPS stored
If
DPS = Yes
and
DPS stored
is
enabled, the latest DPS adjustments
during infusion are saved
automatically at power off for the
next startup.
If
DPS stored
is disabled, the DPS
adjustments must be entered at the
next startup.
Enabled / Disabled
User 5: KVO
KVO1
Sets the
KVO1
flow rate.
0.1
5 mL/h
KVO2
Sets the
KVO2
flow rate.
0.1
5 mL/h
Continuous
Activates the Continuous function.
After the VTBI is completed, the
infusion continues at the
programmed flow rate.
Yes / No
Silence duration
Sets the silence duration alarm in
KVO.
1 min
12 hours
Function
Setting
Description
Range of Settings
Summary of Contents for Agilia SP
Page 139: ...139 17 Spare Parts Catalog Refer to the spare parts catalog of the relevant pump ...
Page 143: ...143 VTBI Volume to Be Infused W Watts Term Description ...
Page 147: ...Local Contacts for Servicing Write your contacts here ...
Page 148: ...11378 2_tm_agilia_sp_eng ...
Page 305: ...Local Contacts for Servicing Write your contacts here ...