84
7.
Disconnect the display board flat cable at [J2].
8.
Disconnect the display board / keypad cable (connection at [J1]).
9.
Remove the 4 screws [6] that fasten the board unit to the upper case.
10.
Remove the display board unit from the upper case.
11.
Disconnect the display board / LCD display connection at [J3].
12.
Unclip and separate the LCD display [8] from the display board [7].
Re-assembling
1.
Perform the dismantling procedure in reverse order to re-assemble the unit.
Pay attention to the direction of the flat cable connections when assembling.
2.
Perform the following calibration using Agilia Partner maintenance software:
LCD contrast
3.
Perform the regular servicing tests. See Quality Control Procedure, page 63.
6 J2 J3
6
6
6
J1
J3
7
8