64
7.12.2
Pre-programming the Pump
You can program the pump before installing the syringe.
1.
Press
to power on the pump.
Syringe installation !!!
is displayed on top of the pump
screen.
2.
Make sure the syringe barrel clasp is folded up against the pump.
The
prog
symbol is displayed.
3.
Press
prog
.
4.
Program the infusion.
5.
Press
exit
to confirm.
6.
When ready, install the syringe.
7.
Press
start
to start the infusion.