![Fresenius Kabi Agilia SP Technical Manual Download Page 28](http://html.mh-extra.com/html/fresenius-kabi/agilia-sp/agilia-sp_technical-manual_2331626028.webp)
28
3
Fundamentals
3.1
Training
3.2
Maintenance Schedule
3.2.1
Preventive Maintenance
3.2.2
Quality Control
The following three training levels outline the specifics needed to maintain and preserve the equipment.
Training level
Description / Prerequisites
1
Training intended for users for on-site maintenance, using the technical documentation of the pump and
specific tools:
Knowledge of mechanics and electronics
Biomedical structure knowledge
2
Training intended for technicians specializing in maintenance performed with specific tools and
procedures:
Good mechanical and electronics knowledge
Two years experience minimum in a biomedical department
3
Training intended for technicians specialized in repair performed in the maintenance department using
specific procedures and tools, including instruments for measurement and adjustments. Complete check-
up using this pump manual:
Good mechanical and electrical knowledge
Good computer knowledge
At least two years experience in a biomedical department
Information
For training, contact your
Fresenius Kabi
sales representative.
Any abnormal functioning or failures must be reported to the qualified technical staff in your organization,
or to your
Fresenius Kabi
representative. In these instances, the pump should not be used.
Warning
In order to maintain the pump’s performance, a preventive maintenance inspection must be carried out at
least once every 3 years. This procedure, which includes changing the battery, should be carried out by
trained and qualified technical personnel in compliance with this document and procedures. Only
authorized personnel should attempt to repair the device.
If these maintenance procedures are not observed, the pump’s correct operation will be impaired.
Upon request by the healthcare facility, a quality control check can be performed on the device every
12 months.
A regular quality check (not included in the guarantee) consists of various inspection operations listed in
this document. See section 6.2, page 40.
Information
These control checks must be performed by trained technical personnel, and are not covered by any
contracts or agreements provided by
Fresenius Kabi
.
For more information, contact your
Fresenius Kabi
sales representative.
Summary of Contents for Agilia SP
Page 139: ...139 17 Spare Parts Catalog Refer to the spare parts catalog of the relevant pump ...
Page 143: ...143 VTBI Volume to Be Infused W Watts Term Description ...
Page 147: ...Local Contacts for Servicing Write your contacts here ...
Page 148: ...11378 2_tm_agilia_sp_eng ...
Page 305: ...Local Contacts for Servicing Write your contacts here ...