![Fresenius Kabi Agilia SP Technical Manual Download Page 52](http://html.mh-extra.com/html/fresenius-kabi/agilia-sp/agilia-sp_technical-manual_2331626052.webp)
52
6.2.12
Test 11: Ageing Test
Test Description
The Ageing test checks that the motor unit is functioning correctly by carrying out
displacements between the 115 mm (4.52 in) and 20 mm (0.79 in) positions.
The motor unit is functioning correctly.
If the test fails, replace the motor unit. See section 8.9, page 102.
Required Tools and Equipment
No specific tools or equipment are required
1.
Access the
Tests
menu. See section 6.2.1, page 40.
2.
Press the arrow keys to select
Test 11: Ageing test
.
3.
Press
enter
.
4.
Press the arrow keys to enter the flow rate to be applied in mL/h.
Available range: 0.1 to 1500 mL/h.
5.
Press
OK
.
6.
Press the arrow keys to enter the motor command mode (
Current
).
Available options:
Low
,
High
,
Maximum
or
Auto
.
7.
Press
OK
.
8.
Press the arrow keys to select the displacement type
Advance
or
Go/return
.
9.
Press
OK
to start the test.
If a battery alarm is present, the test cannot be launched (the
OK
button is not
available).
The pump starts moving the plunger forward.
While an ageing test is in progress, the pump displays the following information:
Flow rate in mL/h
Motor command mode (Current = Low, High, Maximum or Auto)
Animation representing the plunger displacement
The test ends when:
The plunger reaches the 20 mm (0.79 in) position (
Displacement = Advance
).
The plunger reaches the 115 mm (4.52 in) position (
Displacement = Go/return
). The plunger
starts moving backwards when it reaches the 20 mm position.
The user presses the
exit
key.
Information
During the test, the motor rotation is tested and the pump triggers an error if a problem is detected.
At end of test, the pump displays the following information:
Flow rate in mL/h
Motor command mode (
Low, High, Maximum or Auto
)
The test status (Finish)
10.
Press
exit
at end of test to return to the
Test
screen.
Summary of Contents for Agilia SP
Page 139: ...139 17 Spare Parts Catalog Refer to the spare parts catalog of the relevant pump ...
Page 143: ...143 VTBI Volume to Be Infused W Watts Term Description ...
Page 147: ...Local Contacts for Servicing Write your contacts here ...
Page 148: ...11378 2_tm_agilia_sp_eng ...
Page 305: ...Local Contacts for Servicing Write your contacts here ...