29
3.3
Accessing / Exiting the Operating Functions
3.3.1
Accessing the Options Menu
Commands
3.3.2
Exiting the Options Menu
The
Options
menu gives access to the functions for operating the pump. The
Options
menu allows the
operator to customize the pump, edit infusion and maintenance parameters, and run tests.
The
Options
menu comprises the following:
Pump Settings
Basic Profile Configuration / Basic & TCI configuration (for Agilia SP TIVA pumps only)
Profile Menu
Maintenance
When the pump is powered off, simultaneously press both
and
to access the
Options
menu.
Options Menu
Access code?
Description
Pump Settings
No (unless modified by Test20)
Please refer to the IFU.
Profile
Yes
Basic Profile Configuration
Yes
Maintenance
Yes
Information
If the wrong access code is entered,
Error
is displayed.
As long as the pump remains powered on, the code does not need to be entered again.
Operation
Key
Scroll options and change selected value (increasing/decreasing)
Arrow keys
Confirm Value / Move to Next Field
Confirmation key
Select
-
Deselect
Fast Increment key
Go to parent menu (if any) and cancel option
Menu key
Power the pump off and on again to exit the
Options
menu and return to standard operation.
All data, including newly selected and entered option, is saved to memory when the pump is powered off.
Summary of Contents for Agilia SP
Page 139: ...139 17 Spare Parts Catalog Refer to the spare parts catalog of the relevant pump ...
Page 143: ...143 VTBI Volume to Be Infused W Watts Term Description ...
Page 147: ...Local Contacts for Servicing Write your contacts here ...
Page 148: ...11378 2_tm_agilia_sp_eng ...
Page 305: ...Local Contacts for Servicing Write your contacts here ...