6.2.7
Test 6: Displacement
Test Description
The Displacement sensor is a linear potentiometer used to control the linear motion of the
plunger driver and to determine the remaining volume in the syringe.
Run this test to validate the accuracy of the displacement sensor.
The plunger’s driver displacement is correctly calibrated.
This test failed if one of these two conditions is encountered:
The displacement sensor is not calibrated. Calibrate the displacement sensor using
Agilia Partner maintenance software.
The displacement sensor is defective or not correctly connected to the CPU board.
This problem leads to an error code 21:
- Check the cable connection to CPU board (J2 connector).
- Replace the displacement sensor. See section 8.12, page 115.
Required Tools and Equipment
1 displacement verification gauge, 2.56 in (65 mm) long
Prerequisites
The calibration certificate of the displacement verification gauge is valid.
1.
Access the
Tests
menu. See section 6.2.1, page 40.
2.
Press the arrow keys to select
Test 6: Displacement
.
3.
Press
enter
.
4.
Press the arrow keys to scroll through the information screens
.
The following data is displayed:
Conversion
: Sensor value in mV. The displayed voltage value depends on
the plunger head position
Position
: Plunger head position value in mm
Low (20 mm)
: Sensor value in mV when the plunger head is positioned at
20 mm (0.79 in). Value saved during last calibration
High (115 mm)
: Sensor value in mV when the plunger head is positioned
at 115 mm (4.53 in). Value saved during last calibration
Date
: Date of last calibration displayed in the following format:
Month/Day/Year (mm/dd/yyyy)
Number
: Last calibration number
Linearization
: Date of last calibration that included a linearization. This
date is displayed in the following format: Month/Day/Year (mm/dd/yyyy)
CPU BOARD
Displacement
sensor
Main
microprocessor
Linear potentiometer
Interface
Summary of Contents for Agilia SP
Page 139: ...139 17 Spare Parts Catalog Refer to the spare parts catalog of the relevant pump ...
Page 143: ...143 VTBI Volume to Be Infused W Watts Term Description ...
Page 147: ...Local Contacts for Servicing Write your contacts here ...
Page 148: ...11378 2_tm_agilia_sp_eng ...
Page 305: ...Local Contacts for Servicing Write your contacts here ...