6.2.15
Test 20: Pump Code
6.2.16
Test 21: Power Info
Test Description
Run this test to display and modify the code which gives access to the Pump Settings from the
Options
menu. See section 3.3, page 29.
1.
Access the
Tests
menu. See section 6.2.1, page 40.
2.
Press the arrow keys to select
Test 20: Pump code
.
3.
Press
enter
.
The current Pump Settings access code is displayed.
4.
Enter the code
.
5.
Press
OK
to confirm.
6.
Press
exit
to return to the
Tests
menu.
Information
If you do not want to protect access to the Pump settings by a code, enter "0000".
Test Description
Run this test to display power source information on the following parts:
DC voltage
Battery DC voltage charging value
Backup capacitor voltage and temperature
5 V booster voltage.
Information
The backup capacitor voltage and temperature must be systematically checked during the maintenance
protocol.
The AC power board and power supply boards work properly.
DC voltage
out of limits:
Check the connections between power supply board (J5 connector) and AC power
board.
Check the fuse on AC power board.
Replace the power supply board.
Replace the angle bracket supporting the AC power board.
Backup capacitor voltage
and/or
5 V booster voltage
out of limits:
Connect the pump to an AC power supply and wait for 30 minutes.
Check the backup capacitor voltage.
Replace the power supply board if the voltage is outside the limits.
A too high voltage may result in damage to the equipment.
Summary of Contents for Agilia SP
Page 139: ...139 17 Spare Parts Catalog Refer to the spare parts catalog of the relevant pump ...
Page 143: ...143 VTBI Volume to Be Infused W Watts Term Description ...
Page 147: ...Local Contacts for Servicing Write your contacts here ...
Page 148: ...11378 2_tm_agilia_sp_eng ...
Page 305: ...Local Contacts for Servicing Write your contacts here ...