47
6.2.8
Test 7: Syringe Barrel Clasp
Test Description
The Syringe barrel clasp sensor is used to maintain the syringe in the pump, to detect the good
positioning of the syringe and to measure the diameter of the syringe.
Run this test to check:
correct measurement for the syringe diameter
correct functioning of the syringe fastening system
The plunger’s driver displacement is correctly calibrated.
This test failed if one of these two conditions is encountered:
The syringe barrel clasp sensor is not calibrated. Calibrate the syringe barrel clasp
sensor using Agilia Partner maintenance software.
The syringe barrel clasp sensor is defective or not correctly connected to the CPU
board. This problem leads to an error code 20:
- Check the cable connection to CPU board (J1 connector). See figure 2.8,
- Replace the displacement sensor. See section 8.7, page 94.
Required Tools and Equipment
V1 diameter verification gauge, 11.5 mm (0.45 in) diameter
V2 diameter verification gauge, 15.5 mm (0.61 in) diameter
V4 diameter verification gauge, 28 mm (1.10 in) diameter
Prerequisites
The calibration certificates of the diameter verification gauges are valid.
1.
Access the
Tests
menu. See section 6.2.1, page 40.
2.
Press the arrow keys to select
Test 7: Syringe barrel clasp
.
3.
Press
enter
.
CPU BOARD
Syringe barrel
clasp
Main
microprocessor
Opening arms
Interface
Syringe
installation
Optical sensor - disengagement
(plunger head)
Opening levers
Potentiometer
Switch (syringe
detection)
Summary of Contents for Agilia SP
Page 139: ...139 17 Spare Parts Catalog Refer to the spare parts catalog of the relevant pump ...
Page 143: ...143 VTBI Volume to Be Infused W Watts Term Description ...
Page 147: ...Local Contacts for Servicing Write your contacts here ...
Page 148: ...11378 2_tm_agilia_sp_eng ...
Page 305: ...Local Contacts for Servicing Write your contacts here ...