AS-HV Operator’s Manual
30
Doc. 065125-01 10/06
Table 3-5. “Slow” Syringe Speed Settings: Wash Port Dispense
Speed for Concentrator Column/Push or Pull Mode
Speed Setting
Top Speed
(steps/sec)
Flow Rate
(mL/min)
1
40
0.50
2
80
1.00
3
100
1.25
4
120
1.50
5
140
1.75
6
160
2.00
7
180
2.25
8
200
2.50
9
220
2.75
10
240
3.00
Table 3-6. “Slow” Syringe Speed Settings: Sample Aspirate Speed
for Concentrator Column/Pull Mode
Speed Setting
Top Speed
(steps/sec)
Flow Rate
(mL/min)
1
8
0.10
2
16
0.20
3
24
0.30
4
32
0.40
5
40
0.50
6
48
0.60
7
56
0.70
8
64
0.80
9
72
0.90
10
80
1.00